CVE-2021-33886

CVSS V2 Medium 5.8 CVSS V3 High 8.8
Description
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Overview
  • CVE ID
  • CVE-2021-33886
  • Assigner
  • cve@mitre.org
  • Vulnerability Status
  • Analyzed
  • Published Version
  • 2021-08-25T12:15:17
  • Last Modified Date
  • 2022-07-12T17:42:04
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:* 1 OR 012u000062
cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:* 0 OR
cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:* 0 OR
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:A/AC:L/Au:N/C:P/I:P/A:P
  • Access Vector
  • ADJACENT_NETWORK
  • Access Compatibility
  • LOW
  • Authentication
  • NONE
  • Confidentiality Impact
  • PARTIAL
  • Integrity Impact
  • PARTIAL
  • Availability Impact
  • PARTIAL
  • Base Score
  • 5.8
  • Severity
  • MEDIUM
  • Exploitability Score
  • 6.5
  • Impact Score
  • 6.4
CVSS Version 3
  • Version
  • 3.1
  • Vector String
  • CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
  • Attack Vector
  • ADJACENT_NETWORK
  • Attack Compatibility
  • LOW
  • Privileges Required
  • NONE
  • User Interaction
  • NONE
  • Scope
  • UNCHANGED
  • Confidentiality Impact
  • HIGH
  • Availability Impact
  • HIGH
  • Base Score
  • 8.8
  • Base Severity
  • HIGH
  • Exploitability Score
  • 2.8
  • Impact Score
  • 5.9
History
Created Old Value New Value Data Type Notes
2022-05-10 16:08:03 Added to TrackCVE
2022-12-05 09:03:14 2021-08-25T12:15Z 2021-08-25T12:15:17 CVE Published Date updated
2022-12-05 09:03:14 2022-07-12T17:42:04 CVE Modified Date updated
2022-12-05 09:03:14 Analyzed Vulnerability Status updated