CVE-2019-6538
CVSS V2 Low 3.3
CVSS V3 Medium 6.5
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Overview
- CVE ID
- CVE-2019-6538
- Assigner
- ics-cert@hq.dhs.gov
- Vulnerability Status
- Analyzed
- Published Version
- 2019-03-25T22:29:00
- Last Modified Date
- 2020-10-06T13:19:18
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24950:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:o:medtronic:mycarelink_monitor_firmware:24952:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:mycarelink_monitor:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:carelink_monitor_firmware:2490c:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:carelink_monitor:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:carelink_2090_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:carelink_2090:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:amplia_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:amplia_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:claria_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:claria_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:compia_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:compia_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:concerto_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:concerto_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:concerto_ii_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:concerto_ii_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:consulta_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:consulta_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:evera_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:evera_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:maximo_ii_crt-d_and_lcd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:maximo_ii_crt-d_and_lcd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:mirro_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:mirro_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:nayamed_nd_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:nayamed_nd_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:primo_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:primo_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:protecta_icd_and_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:protecta_icd_and_crt-d:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:secura_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:secura_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:virtuoso_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:virtuoso_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:virtuoso_ii_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:virtuoso_ii_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:visia_af_icd_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:visia_af_icd:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronic:viva_crt-d_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronic:viva_crt-d:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:A/AC:L/Au:N/C:N/I:P/A:N
- Access Vector
- ADJACENT_NETWORK
- Access Compatibility
- LOW
- Authentication
- NONE
- Confidentiality Impact
- NONE
- Integrity Impact
- PARTIAL
- Availability Impact
- NONE
- Base Score
- 3.3
- Severity
- LOW
- Exploitability Score
- 6.5
- Impact Score
- 2.9
CVSS Version 3
- Version
- 3.1
- Vector String
- CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
- Attack Vector
- ADJACENT_NETWORK
- Attack Compatibility
- LOW
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality Impact
- NONE
- Availability Impact
- NONE
- Base Score
- 6.5
- Base Severity
- MEDIUM
- Exploitability Score
- 2.8
- Impact Score
- 3.6
References
Reference URL | Reference Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01 | Mitigation Third Party Advisory US Government Resource |
http://www.securityfocus.com/bid/107544 | Third Party Advisory VDB Entry |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2019-6538 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2019-6538 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 17:10:16 | Added to TrackCVE | |||
2022-12-03 18:38:05 | 2019-03-25T22:29Z | 2019-03-25T22:29:00 | CVE Published Date | updated |
2022-12-03 18:38:05 | 2020-10-06T13:19:18 | CVE Modified Date | updated | |
2022-12-03 18:38:05 | Analyzed | Vulnerability Status | updated |