Description
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Overview
  • CVE ID
  • CVE-2018-14781
  • Assigner
  • ics-cert@hq.dhs.gov
  • Vulnerability Status
  • Modified
  • Published Version
  • 2018-08-13T21:48:01
  • Last Modified Date
  • 2019-10-09T23:35:11
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:* 0 OR
AND
cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:* 0 OR
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:A/AC:M/Au:N/C:N/I:P/A:N
  • Access Vector
  • ADJACENT_NETWORK
  • Access Compatibility
  • MEDIUM
  • Authentication
  • NONE
  • Confidentiality Impact
  • NONE
  • Integrity Impact
  • PARTIAL
  • Availability Impact
  • NONE
  • Base Score
  • 2.9
  • Severity
  • LOW
  • Exploitability Score
  • 5.5
  • Impact Score
  • 2.9
CVSS Version 3
  • Version
  • 3.0
  • Vector String
  • CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
  • Attack Vector
  • ADJACENT_NETWORK
  • Attack Compatibility
  • HIGH
  • Privileges Required
  • NONE
  • User Interaction
  • NONE
  • Scope
  • UNCHANGED
  • Confidentiality Impact
  • NONE
  • Availability Impact
  • NONE
  • Base Score
  • 5.3
  • Base Severity
  • MEDIUM
  • Exploitability Score
  • 1.6
  • Impact Score
  • 3.6
References
Reference URL Reference Tags
http://www.securityfocus.com/bid/105044 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 Third Party Advisory US Government Resource
History
Created Old Value New Value Data Type Notes
2022-05-10 17:06:20 Added to TrackCVE
2022-12-03 11:06:28 2018-08-13T21:48Z 2018-08-13T21:48:01 CVE Published Date updated
2022-12-03 11:06:28 2019-10-09T23:35:11 CVE Modified Date updated
2022-12-03 11:06:28 Modified Vulnerability Status updated