CVE-2018-14781
CVSS V2 Low 2.9
CVSS V3 Medium 5.3
Description
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Overview
- CVE ID
- CVE-2018-14781
- Assigner
- ics-cert@hq.dhs.gov
- Vulnerability Status
- Modified
- Published Version
- 2018-08-13T21:48:01
- Last Modified Date
- 2019-10-09T23:35:11
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:medtronicdiabetes:508_minimed_insulin_pump_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:508_minimed_insulin_pump:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:522_paradigm_real-time_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:522_paradigm_real-time:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:722_paradigm_real-time_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:722_paradigm_real-time:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:523_paradigm_revel_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:523_paradigm_revel:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:723_paradigm_revel_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:723_paradigm_revel:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:523k_paradigm_revel_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:523k_paradigm_revel:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:723k_paradigm_revel_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:723k_paradigm_revel:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:551_minimed_530g_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:551_minimed_530g:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:medtronicdiabetes:751_minimed_530g_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:medtronicdiabetes:751_minimed_530g:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:A/AC:M/Au:N/C:N/I:P/A:N
- Access Vector
- ADJACENT_NETWORK
- Access Compatibility
- MEDIUM
- Authentication
- NONE
- Confidentiality Impact
- NONE
- Integrity Impact
- PARTIAL
- Availability Impact
- NONE
- Base Score
- 2.9
- Severity
- LOW
- Exploitability Score
- 5.5
- Impact Score
- 2.9
CVSS Version 3
- Version
- 3.0
- Vector String
- CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
- Attack Vector
- ADJACENT_NETWORK
- Attack Compatibility
- HIGH
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality Impact
- NONE
- Availability Impact
- NONE
- Base Score
- 5.3
- Base Severity
- MEDIUM
- Exploitability Score
- 1.6
- Impact Score
- 3.6
References
Reference URL | Reference Tags |
---|---|
http://www.securityfocus.com/bid/105044 | Third Party Advisory VDB Entry |
https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02 | Third Party Advisory US Government Resource |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2018-14781 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2018-14781 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 17:06:20 | Added to TrackCVE | |||
2022-12-03 11:06:28 | 2018-08-13T21:48Z | 2018-08-13T21:48:01 | CVE Published Date | updated |
2022-12-03 11:06:28 | 2019-10-09T23:35:11 | CVE Modified Date | updated | |
2022-12-03 11:06:28 | Modified | Vulnerability Status | updated |