CVE-2018-10601
CVSS V2 Medium 5.4
CVSS V3 High 8.2
Description
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Overview
- CVE ID
- CVE-2018-10601
- Assigner
- ics-cert@hq.dhs.gov
- Vulnerability Status
- Analyzed
- Published Version
- 2018-06-05T20:29:00
- Last Modified Date
- 2021-05-10T15:08:11
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:philips:intellivue_mp2_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mp2:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_x2_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_x2:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mp30_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mp30:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mp50_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mp50:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mp70_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mp70:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_np90_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_np90:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx700_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx700:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx800_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx800:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx400_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx400:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx450_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx450:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx500_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx500:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx550_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx550:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_x3_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_x3:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:intellivue_mx100_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:intellivue_mx100:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm20_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm20:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm30_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm30:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm40_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm40:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:philips:avalon_fetal\/maternal_monitors_fm50_firmware:-:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:philips:avalon_fetal\/maternal_monitors_fm50:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:A/AC:M/Au:N/C:P/I:P/A:P
- Access Vector
- ADJACENT_NETWORK
- Access Compatibility
- MEDIUM
- Authentication
- NONE
- Confidentiality Impact
- PARTIAL
- Integrity Impact
- PARTIAL
- Availability Impact
- PARTIAL
- Base Score
- 5.4
- Severity
- MEDIUM
- Exploitability Score
- 5.5
- Impact Score
- 6.4
CVSS Version 3
- Version
- 3.1
- Vector String
- CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H
- Attack Vector
- ADJACENT_NETWORK
- Attack Compatibility
- HIGH
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- CHANGED
- Confidentiality Impact
- HIGH
- Availability Impact
- HIGH
- Base Score
- 8.2
- Base Severity
- HIGH
- Exploitability Score
- 1.6
- Impact Score
- 6
References
Reference URL | Reference Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01 | Third Party Advisory US Government Resource |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2018-10601 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2018-10601 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 16:16:35 | Added to TrackCVE | |||
2022-12-03 07:13:57 | 2018-06-05T20:29Z | 2018-06-05T20:29:00 | CVE Published Date | updated |
2022-12-03 07:13:57 | 2021-05-10T15:08:11 | CVE Modified Date | updated | |
2022-12-03 07:13:57 | Analyzed | Vulnerability Status | updated |