CVE-2017-9658

CVSS V2 Medium 6.1 CVSS V3 Medium 6.5
Description
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Overview
  • CVE ID
  • CVE-2017-9658
  • Assigner
  • ics-cert@hq.dhs.gov
  • Vulnerability Status
  • Modified
  • Published Version
  • 2018-04-30T15:29:00
  • Last Modified Date
  • 2019-10-09T23:30:47
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:* 1 OR b.06.18
cpe:2.3:h:philips:intellivue_mx40:-:*:*:*:*:*:*:* 0 OR
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:A/AC:L/Au:N/C:N/I:N/A:C
  • Access Vector
  • ADJACENT_NETWORK
  • Access Compatibility
  • LOW
  • Authentication
  • NONE
  • Confidentiality Impact
  • NONE
  • Integrity Impact
  • NONE
  • Availability Impact
  • COMPLETE
  • Base Score
  • 6.1
  • Severity
  • MEDIUM
  • Exploitability Score
  • 6.5
  • Impact Score
  • 6.9
CVSS Version 3
  • Version
  • 3.0
  • Vector String
  • CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
  • Attack Vector
  • ADJACENT_NETWORK
  • Attack Compatibility
  • LOW
  • Privileges Required
  • NONE
  • User Interaction
  • NONE
  • Scope
  • UNCHANGED
  • Confidentiality Impact
  • NONE
  • Availability Impact
  • HIGH
  • Base Score
  • 6.5
  • Base Severity
  • MEDIUM
  • Exploitability Score
  • 2.8
  • Impact Score
  • 3.6
References
Reference URL Reference Tags
https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Third Party Advisory US Government Resource
http://www.securityfocus.com/bid/100813 Third Party Advisory VDB Entry
History
Created Old Value New Value Data Type Notes
2022-05-10 17:08:15 Added to TrackCVE
2022-12-03 05:46:27 2018-04-30T15:29Z 2018-04-30T15:29:00 CVE Published Date updated
2022-12-03 05:46:27 2019-10-09T23:30:47 CVE Modified Date updated
2022-12-03 05:46:27 Modified Vulnerability Status updated