CVE-2017-12725
CVSS V2 Medium 6.8
CVSS V3 Medium 5.6
Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Overview
- CVE ID
- CVE-2017-12725
- Assigner
- ics-cert@hq.dhs.gov
- Vulnerability Status
- Analyzed
- Published Version
- 2018-02-15T10:29:00
- Last Modified Date
- 2018-03-02T14:40:15
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:* | 1 | OR | ||
cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:N/AC:M/Au:N/C:P/I:P/A:P
- Access Vector
- NETWORK
- Access Compatibility
- MEDIUM
- Authentication
- NONE
- Confidentiality Impact
- PARTIAL
- Integrity Impact
- PARTIAL
- Availability Impact
- PARTIAL
- Base Score
- 6.8
- Severity
- MEDIUM
- Exploitability Score
- 8.6
- Impact Score
- 6.4
CVSS Version 3
- Version
- 3.0
- Vector String
- CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
- Attack Vector
- NETWORK
- Attack Compatibility
- HIGH
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality Impact
- LOW
- Availability Impact
- LOW
- Base Score
- 5.6
- Base Severity
- MEDIUM
- Exploitability Score
- 2.2
- Impact Score
- 3.4
References
Reference URL | Reference Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A | Third Party Advisory US Government Resource |
http://www.securityfocus.com/bid/100665 | Third Party Advisory VDB Entry |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2017-12725 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2017-12725 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 18:47:59 | Added to TrackCVE | |||
2022-12-03 02:11:24 | 2018-02-15T10:29Z | 2018-02-15T10:29:00 | CVE Published Date | updated |
2022-12-03 02:11:24 | 2018-03-02T14:40:15 | CVE Modified Date | updated | |
2022-12-03 02:11:24 | Analyzed | Vulnerability Status | updated |