CVE-2017-12718

CVSS V2 Medium 6.8 CVSS V3 High 8.1
Description
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Overview
  • CVE ID
  • CVE-2017-12718
  • Assigner
  • ics-cert@hq.dhs.gov
  • Vulnerability Status
  • Modified
  • Published Version
  • 2018-02-15T10:29:00
  • Last Modified Date
  • 2019-10-09T23:23:11
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.1:*:*:*:*:*:*:* 1 OR
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.5:*:*:*:*:*:*:* 1 OR
cpe:2.3:o:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:1.6:*:*:*:*:*:*:* 1 OR
cpe:2.3:h:smiths-medical:medfusion_4000_wireless_syringe_infusion_pump:-:*:*:*:*:*:*:* 0 OR
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:N/AC:M/Au:N/C:P/I:P/A:P
  • Access Vector
  • NETWORK
  • Access Compatibility
  • MEDIUM
  • Authentication
  • NONE
  • Confidentiality Impact
  • PARTIAL
  • Integrity Impact
  • PARTIAL
  • Availability Impact
  • PARTIAL
  • Base Score
  • 6.8
  • Severity
  • MEDIUM
  • Exploitability Score
  • 8.6
  • Impact Score
  • 6.4
CVSS Version 3
  • Version
  • 3.0
  • Vector String
  • CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
  • Attack Vector
  • NETWORK
  • Attack Compatibility
  • HIGH
  • Privileges Required
  • NONE
  • User Interaction
  • NONE
  • Scope
  • UNCHANGED
  • Confidentiality Impact
  • HIGH
  • Availability Impact
  • HIGH
  • Base Score
  • 8.1
  • Base Severity
  • HIGH
  • Exploitability Score
  • 2.2
  • Impact Score
  • 5.9
References
Reference URL Reference Tags
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A Third Party Advisory US Government Resource
https://www.exploit-db.com/exploits/43776/ Third Party Advisory VDB Entry
http://www.securityfocus.com/bid/101252 Third Party Advisory VDB Entry
http://www.securityfocus.com/bid/100665 Third Party Advisory VDB Entry
History
Created Old Value New Value Data Type Notes
2022-05-10 17:11:41 Added to TrackCVE
2022-12-03 02:11:04 2018-02-15T10:29Z 2018-02-15T10:29:00 CVE Published Date updated
2022-12-03 02:11:04 2019-10-09T23:23:11 CVE Modified Date updated
2022-12-03 02:11:04 Modified Vulnerability Status updated