CVE-2017-12712
CVSS V2 High 8.3
CVSS V3 High 8.8
Description
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Overview
- CVE ID
- CVE-2017-12712
- Assigner
- ics-cert@hq.dhs.gov
- Vulnerability Status
- Modified
- Published Version
- 2018-04-25T13:29:00
- Last Modified Date
- 2019-10-09T23:23:11
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:* | 1 | OR | f0b.0e.7e | |
cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:* | 1 | OR | f0b.0e.7e | |
cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:* | 1 | OR | f10.08.6c | |
cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:* | 1 | OR | f10.08.6c | |
cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:assurity_firmware:*:*:*:*:*:*:*:* | 1 | OR | f14.07.80 | |
cpe:2.3:h:abbott:assurity:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:allure_firmware:*:*:*:*:*:*:*:* | 1 | OR | f14.07.80 | |
cpe:2.3:h:abbott:allure:-:*:*:*:*:*:*:* | 0 | OR | ||
AND | ||||
cpe:2.3:o:abbott:assurity_mri_firmware:*:*:*:*:*:*:*:* | 1 | OR | f17.01.49 | |
cpe:2.3:h:abbott:assurity_mri:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:A/AC:L/Au:N/C:C/I:C/A:C
- Access Vector
- ADJACENT_NETWORK
- Access Compatibility
- LOW
- Authentication
- NONE
- Confidentiality Impact
- COMPLETE
- Integrity Impact
- COMPLETE
- Availability Impact
- COMPLETE
- Base Score
- 8.3
- Severity
- HIGH
- Exploitability Score
- 6.5
- Impact Score
- 10
CVSS Version 3
- Version
- 3.0
- Vector String
- CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Attack Vector
- ADJACENT_NETWORK
- Attack Compatibility
- LOW
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality Impact
- HIGH
- Availability Impact
- HIGH
- Base Score
- 8.8
- Base Severity
- HIGH
- Exploitability Score
- 2.8
- Impact Score
- 5.9
References
Reference URL | Reference Tags |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 | Third Party Advisory US Government Resource |
http://www.securityfocus.com/bid/100523 | Third Party Advisory VDB Entry |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2017-12712 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2017-12712 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 17:11:42 | Added to TrackCVE | |||
2022-12-03 05:39:07 | 2018-04-25T13:29Z | 2018-04-25T13:29:00 | CVE Published Date | updated |
2022-12-03 05:39:07 | 2019-10-09T23:23:11 | CVE Modified Date | updated | |
2022-12-03 05:39:07 | Modified | Vulnerability Status | updated |