CVE-2017-11578
CVSS V2 Medium 4.3
CVSS V3 Medium 5.9
Description
It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.
Overview
- CVE ID
- CVE-2017-11578
- Assigner
- cve@mitre.org
- Vulnerability Status
- Analyzed
- Published Version
- 2019-07-02T21:15:09
- Last Modified Date
- 2019-07-15T13:10:30
Weakness Enumerations
CPE Configuration (Product)
CPE | Vulnerable | Operator | Version Start | Version End |
---|---|---|---|---|
AND | ||||
cpe:2.3:o:blipcare:wi-fi_blood_pressure_monitor_firmware:*:*:*:*:*:*:*:* | 1 | OR | bp700_10.1 | |
cpe:2.3:h:blipcare:wi-fi_blood_pressure_monitor:-:*:*:*:*:*:*:* | 0 | OR |
CVSS Version 2
- Version
- 2.0
- Vector String
- AV:N/AC:M/Au:N/C:P/I:N/A:N
- Access Vector
- NETWORK
- Access Compatibility
- MEDIUM
- Authentication
- NONE
- Confidentiality Impact
- PARTIAL
- Integrity Impact
- NONE
- Availability Impact
- NONE
- Base Score
- 4.3
- Severity
- MEDIUM
- Exploitability Score
- 8.6
- Impact Score
- 2.9
CVSS Version 3
- Version
- 3.0
- Vector String
- CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
- Attack Vector
- NETWORK
- Attack Compatibility
- HIGH
- Privileges Required
- NONE
- User Interaction
- NONE
- Scope
- UNCHANGED
- Confidentiality Impact
- HIGH
- Availability Impact
- NONE
- Base Score
- 5.9
- Base Severity
- MEDIUM
- Exploitability Score
- 2.2
- Impact Score
- 3.6
References
Reference URL | Reference Tags |
---|---|
https://github.com/ethanhunnt/IoT_vulnerabilities/blob/master/Blipcare_sec_issues.pdf | Exploit Third Party Advisory |
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html | Third Party Advisory VDB Entry |
https://seclists.org/bugtraq/2019/Jun/8 | Mailing List Third Party Advisory |
Sources
Source Name | Source URL |
---|---|
NIST | https://nvd.nist.gov/vuln/detail/CVE-2017-11578 |
MITRE | https://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2017-11578 |
History
Created | Old Value | New Value | Data Type | Notes |
---|---|---|---|---|
2022-05-10 17:41:06 | Added to TrackCVE | |||
2022-12-03 22:46:10 | 2019-07-02T21:15Z | 2019-07-02T21:15:09 | CVE Published Date | updated |
2022-12-03 22:46:10 | 2019-07-15T13:10:30 | CVE Modified Date | updated | |
2022-12-03 22:46:10 | Analyzed | Vulnerability Status | updated |