CVE-2014-5406

CVSS V2 High 9.3 CVSS V3 None
Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Overview
  • CVE ID
  • CVE-2014-5406
  • Assigner
  • ics-cert@hq.dhs.gov
  • Vulnerability Status
  • Analyzed
  • Published Version
  • 2015-07-06T19:59:00
  • Last Modified Date
  • 2015-07-08T15:18:06
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:* 1 OR 5.0
cpe:2.3:h:hospira:lifecare_pca3:-:*:*:*:*:*:*:* 0 OR
cpe:2.3:h:hospira:lifecare_pca5:-:*:*:*:*:*:*:* 0 OR
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:N/AC:M/Au:N/C:C/I:C/A:C
  • Access Vector
  • NETWORK
  • Access Compatibility
  • MEDIUM
  • Authentication
  • NONE
  • Confidentiality Impact
  • COMPLETE
  • Integrity Impact
  • COMPLETE
  • Availability Impact
  • COMPLETE
  • Base Score
  • 9.3
  • Severity
  • HIGH
  • Exploitability Score
  • 8.6
  • Impact Score
  • 10
History
Created Old Value New Value Data Type Notes
2022-05-10 10:22:47 Added to TrackCVE
2022-12-02 05:22:23 2015-07-06T19:59Z 2015-07-06T19:59:00 CVE Published Date updated
2022-12-02 05:22:23 2015-07-08T15:18:06 CVE Modified Date updated
2022-12-02 05:22:23 Analyzed Vulnerability Status updated