Description
Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 components, and Xper Information Management servers and workstations for Flex Cardio products before XperConnect 1.5.4.053 SP2 allows remote attackers to execute arbitrary code via a crafted HTTP request to the Connect broker on TCP port 6000.
Overview
  • CVE ID
  • CVE-2013-2808
  • Assigner
  • ics-cert@hq.dhs.gov
  • Vulnerability Status
  • Analyzed
  • Published Version
  • 2013-10-05T10:55:03
  • Last Modified Date
  • 2013-10-07T17:56:44
CPE Configuration (Product)
CPE Vulnerable Operator Version Start Version End
AND
cpe:2.3:a:philips:xper_information_management_physiomonitoring_5:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:a:philips:xperconnect:*:*:*:*:*:*:*:* 1 OR 1.5.4.053
AND
cpe:2.3:a:philips:xper_information_management_vascular_monitoring_5:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:a:philips:xperconnect:*:*:*:*:*:*:*:* 1 OR 1.5.4.053
AND
cpe:2.3:h:philips:xper_flex_cardio:-:*:*:*:*:*:*:* 1 OR
cpe:2.3:a:philips:xperconnect:*:*:*:*:*:*:*:* 1 OR 1.5.4.053
CVSS Version 2
  • Version
  • 2.0
  • Vector String
  • AV:N/AC:M/Au:N/C:C/I:C/A:C
  • Access Vector
  • NETWORK
  • Access Compatibility
  • MEDIUM
  • Authentication
  • NONE
  • Confidentiality Impact
  • COMPLETE
  • Integrity Impact
  • COMPLETE
  • Availability Impact
  • COMPLETE
  • Base Score
  • 9.3
  • Severity
  • HIGH
  • Exploitability Score
  • 8.6
  • Impact Score
  • 10
References
Reference URL Reference Tags
http://ics-cert.us-cert.gov/advisories/ICSA-13-277-01 US Government Resource
History
Created Old Value New Value Data Type Notes
2022-05-10 10:42:05 Added to TrackCVE
2022-12-01 20:02:30 2013-10-05T10:55Z 2013-10-05T10:55:03 CVE Published Date updated
2022-12-01 20:02:30 2013-10-07T17:56:44 CVE Modified Date updated
2022-12-01 20:02:30 Analyzed Vulnerability Status updated